|
References  |
|
|
| *references marked with an asterisk below are user-added references |
|
U.S. References |
|
|
| Add a new US reference: |
| | Reference | Relevancy | Comments | Reference | Relevancy | Comments | 2930377
|      Your vote accepted [0 after 0 votes] | | 3435824
|      Your vote accepted [0 after 0 votes] | | 3769981
|      Your vote accepted [0 after 0 votes] | | 4877031 Conway 606/194 Oct,1989 |      Your vote accepted [0 after 0 votes] | | 4832028 Patel 606/194 May,1989 |      Your vote accepted [0 after 0 votes] | | 4820271 Deutsch 604/99.01 Apr,1989 |      Your vote accepted [0 after 0 votes] | | 4790315 Mueller, Jr. 606/194 Dec,1988 |      Your vote accepted [0 after 0 votes] | | 4771777 Horzewski 606/194 Sep,1988 |      Your vote accepted [0 after 0 votes] | | 4763654 Jang 606/195 Aug,1988 |      Your vote accepted [0 after 0 votes] | | 4762129 Bonzel 606/194 Aug,1988 |      Your vote accepted [0 after 0 votes] | | 4748982 Horzewski 606/192 Jun,1988 |      Your vote accepted [0 after 0 votes] | | 4661094 Simpson 604/8 Apr,1987 |      Your vote accepted [0 after 0 votes] | | 4646742 Packard 606/194 Mar,1987 |      Your vote accepted [0 after 0 votes] | | 4597755 Samson 604/103.09 Jul,1986 |      Your vote accepted [0 after 0 votes] | | 4581017 Sahota 604/101.01 Apr,1986 |      Your vote accepted [0 after 0 votes] | | 4447227 Kotsanis 604/95.03 May,1984 |      Your vote accepted [0 after 0 votes] | | 4423725 Baran 128/207.15 Jan,1984 |      Your vote accepted [0 after 0 votes] | | 4329993 Lieber 604/98.01 May,1982 |      Your vote accepted [0 after 0 votes] | | 4230119 Blum 606/194 Oct,1980 |      Your vote accepted [0 after 0 votes] | | 4198981 Sinnreich 606/193 Apr,1980 |      Your vote accepted [0 after 0 votes] | | 4195637 Gruntzig 604/509 Apr,1980 |      Your vote accepted [0 after 0 votes] | | 3889686 Duturbure 604/102.03 Jun,1975 |      Your vote accepted [0 after 0 votes] | | 3889685 Miller, Jr. 604/8 Jun,1975 |      Your vote accepted [0 after 0 votes] | | | | | |
|
|
|
|
U.S. References |
|
|
Foreign References |
|
|
|
|
|
|
Foreign References |
|
|
Other References |
|
|
|
|
|
|
Other References |
|
|
|
|
|
References  |
|
|
Claims  |
|
|
What is claimed is:
1. A perfusion balloon catheter for use in combination with a guide wire to
dilate a stenosed coronary artery, the perfusion balloon catheter
comprising:
an elongate shaft having an inflation lumen and a guide wire lumen integral
therewith, with the guide wire lumen having a generally bevelled opening
at its distal end;
a separate balloon head connected to a distal end of the shaft, the balloon
head defining a relatively short blood flow channel, an open proximal end
of the blood flow channel being defined by a generally bevelled opening
which is opposite to, generally axially aligned with, and relatively
closely longitudinally spaced from the bevelled opening at the distal end
of the guide wire lumen, and the blood flow channel having a larger
cross-section than the guide wire lumen;
a balloon mounted around the balloon head and defining a balloon interior
which is in communication with a distal end of the inflation lumen; and
a distal tip connected to and extending distally from the balloon, the
distal tip having a central passage aligned with the blood flow channel to
permit blood flow from the artery into the bevelled opening of the blood
flow channel, through the blood flow channel and through the central
passage while the balloon is inflated, and also to permit the guide wire
to move through the bevelled opening of the blood flow channel, through
the channel and through the central passage.
2. The catheter of claim 1 and further comprising:
a manifold connected to a proximal end of the shaft, the manifold including
a guide wire port in communication with the guide wire lumen and an
inflation port in communication with the inflation lumen.
3. The catheter of claim 2 wherein the manifold further includes a through
port in communication with the guide wire lumen.
4. The catheter of claim 1 wherein the inflation lumen and the guide wire
lumen are positioned side-by-side.
5. The catheter of claim 1 wherein the distal tip is generally tapered with
a smaller diameter at its distal end than at its proximal end.
6. The catheter of claim 5 wherein the distal tip has a greater flexibility
than the balloon head.
7. The catheter of claim 1 wherein the balloon and the distal tip are an
integral polymeric material.
8. The catheter of claim 1 wherein the balloon has a distal segment bonded
to a distal portion of the balloon head.
9. The catheter of claim 8 wherein the balloon has a proximal segment
bonded to a distal portion of the shaft and a proximal portion of the
balloon head.
10. The catheter of claim 9 wherein the distal portion of the shaft
contains only the inflation lumen.
11. The catheter of claim 10 wherein the distal portion of the shaft has a
reduced cross-section with respect to other more proximal portions of the
shaft.
12. The catheter of claim 11 wherein the proximal portion of the balloon
head has a reduced outer dimension with respect to more distal portions.
13. The catheter of claim 12 wherein the proximal portion of the balloon
head and the distal portion of the shaft have mating surfaces which are
bonded together.
14. The catheter of claim 1 wherein the balloon head comprises:
a polymeric tube which has the blood flow channel extending therethrough;
and
reinforcing means for reinforcing the tube to prevent collapse of the tube
which would restrict blood flow through the blood-flow channel during
inflation of the balloon.
15. The catheter of claim 14 wherein the reinforcing means comprises a
helically wound coil bonded to the tube.
16. The catheter of claim 15 wherein the coil is formed by a flat metal
ribbon.
17. The catheter of claim 15 wherein the coil is bonded to an interior
surface of the tube.
18. The catheter of claim 14 and further comprising a low friction coating
on an interior surface of the tube.
19. The catheter of claim 1 wherein the distal tip further includes a side
port therein.
20. The catheter of claim 1 and further comprising:
means positioned adjacent the open distal end of the guide wire lumen for
guiding a guide wire toward the open proximal end of the blood flow
channel.
21. The catheter of claim 20 wherein the means positioned adjacent the open
distal end comprises a resilient flap attached to the shaft.
22. The catheter of claim 1 wherein a leading edge of the bevelled opening
of the blood flow channel extends into and overlaps a leading edge of the
bevelled opening of the guide wire lumen.
23. The catheter of claim 1 wherein the blood flow channel is only slightly
longer than the balloon.
24. A perfusion balloon catheter for use in combination with a guide wire
to dilate a stenosed coronary artery, the perfusion balloon catheter
comprising:
an elongate shaft having an inflation lumen and a guide wire lumen integral
therewith;
a separate balloon head connected to a distal end of the shaft, the balloon
head defining a relatively short blood flow channel, an open proximal end
of the blood flow channel being generally axially aligned with and
relatively closely longitudinally spaced from an open distal end of the
guide wire lumen, the balloon head including a polymeric tube which has
the blood flow channel extending therethrough and reinforcing means for
reinforcing the tube to prevent collapse of the tube, and with the blood
flow channel having a larger cross-section than the guide wire lumen; and
a balloon mounted around the balloon head and defining a balloon interior
which is in communication with the distal end of the inflation lumen.
25. The catheter of claim 24 wherein the reinforcing means comprises a
helically wound coil bonded to the tube.
26. The catheter of claim 25 wherein the coil is formed by a flat metal
ribbon.
27. The catheter of claim 25 wherein the coil is bonded to an interior
surface of the tube.
28. The catheter of claim 25 wherein the coil is formed by a radiopaque
metal.
29. The catheter of claim 24 and further comprising a low friction coating
on an interior surface of the tube.
30. The catheter of claim 24 wherein the balloon has a distal segment
bonded to a distal portion of the balloon head.
31. The catheter of claim 30 wherein the balloon has a proximal segment
bonded to a distal portion of the shaft and a proximal portion of the
balloon head.
32. The catheter of claim 31 wherein the distal portion of the shaft
contains only the inflation lumen.
33. The catheter of claim 32 wherein the distal portion of the shaft has a
reduced cross-section with respect to other more proximal portions of the
shaft.
34. The catheter of claim 33 wherein the proximal portion of the balloon
head has a reduced outer dimension with respect to more distal portions.
35. The catheter of claim 34 wherein the proximal portion of the balloon
head and the distal portion of the shaft have mating surfaces which are
bonded together.
36. The catheter of claim 24 wherein the balloon head has a bevelled
opening at its proximal end.
37. The catheter of claim 24 wherein the guide wire lumen has a bevelled
opening at its distal end.
38. The catheter of claim 24 and further comprising:
means positioned adjacent the open distal end of the guide wire lumen for
guiding a guide wire toward the open proximal end of the blood flow
channel.
39. The catheter of claim 38 wherein the means positioned adjacent the open
distal end comprises a resilient flap attached to the shaft.
40. The catheter of claim 24 wherein a leading edge of the open proximal
end of the balloon head extends into and overlaps a leading edge of the
open distal end of the guide wire lumen.
41. The catheter of claim 24 wherein the blood flow channel is only
slightly longer than the balloon.
42. A method of performing angioplasty, the method comprising:
positioning a guide wire within a patient's body so that the guide wire
crosses a stenosis;
advancing a passive perfusion balloon catheter over the guide wire until a
balloon of the catheter is positioned in the stenosis, the catheter
including a balloon head on which the balloon is mounted and which
provides a relatively short blood flow channel through which the guide
wire extends, and a separate, relatively long shaft on which the balloon
head is mounted which includes an inflation lumen connected to the balloon
and an internal guide wire lumen through which the guide wire extends,
with the relatively short blood flow channel having a larger cross-section
than the guide wire lumen;
partially withdrawing the guide wire so that the guide wire is in the guide
wire lumen but is not in the blood flow channel to permit perfusion of
blood from the artery through the blood flow channel in a proximal to
distal direction; and inflating the balloon.
43. The method of claim 42 and further comprising:
readvancing the guide wire from the guide wire lumen through the blood flow
channel so that the guide wire once again extends across the stenosis;
deflating the balloon; and
withdrawing the catheter while leaving the guide wire in position in the
patient's body.
44. The method of claim 42, and further comprising the steps of:
readvancing the guide wire from the guide wire lumen through the blood flow
channel; and
positively guiding a distal end of the guide wire into a proximal end of
the blood flow channel as the guide wire is re-advanced. |
|
|
|
|
Claims  |
|
|
Description  |
|
|
BACKGROUND OF THE INVENTION
1. Field of the Invention.
The present invention relates to the field of angioplasty. In particular,
the present invention relates to a balloon catheter which provides
prolonged dilatations without blocking blood flow by use of passive
perfusion.
2. Description of the Prior Art.
Angioplasty has gained wide acceptance in recent years as an efficient and
effective method for treating types of vascular diseases. In particular,
angioplasty is widely used for opening stenoses in coronary arteries,
although it is also used for treatment of stenoses in other parts of the
vascular system. The most widely used form of angioplasty is called
percutaneous transluminal coronary angioplasty (PTCA) and makes use of a
dilatation catheter which has an inflatable balloon at its distal end.
Using fluoroscopy, the physician guides the catheter through the vascular
system until the balloon is positioned across the stenosis. The balloon is
then inflated by supplying fluid under pressure through an inflation lumen
to the balloon. The inflation of the balloon opens the artery to
reestablish acceptable blood flow through the artery.
An initial concern about PTCA was the temporary blockage of blood flow
during balloon inflation. As cardiologists gained clinical experience with
PTCA, the concern with temporary blood flow blockage declined; the vast
majority of patients tolerate 30 to 60 second dilatations quite well.
Concurrently, cardiologists discovered that prolonged dilatations can help
overcome certain kinds of complications encountered with the angioplasty.
Prolonged dilatations of several minutes are used to deal with
dissections, intimal flaps, acute thrombosis and vessel spasms. The
profound ischemia of a long dilatation is outweighed by the potential
prevention of emergency coronary bypass surgery.
In the prior art, methods for enabling prolonged dilatations have been
cumbersome, have been experimental, or have had potentially harmful side
effects. Consequently, there has been no definitive study of the effects
of prolonged dilations on the efficacy of PTCA.
In order to perform prolonged dilatations, several approaches have been
suggested. These include the use of pharmacologic agents to improve
myocardial tolerance of ischemia, synchronized retroprofusion, mechnical
pump distal perfusion, and auto or passive perfusion.
The use of pharmacologic agents treats the symptoms of ischemia, but not
the cause. As a result, this approach is inherently limited.
Synchronized retroprofusion involves pumping blood during diastole into the
coronary sinus and then subselectively into the regional coronary veins
which drain the jeopardized myocardium. This approach potentially offers
nearly complete myocardial perfusion. The disadvantage of synchronized
retroprofusion, however, is that it is complicated and cumbersome.
Mechanical pump distal perfusion involves pumping blood (or other
perfusate) through a lumen of the PTCA catheter. The need to pump through
the PTCA catheter requires some form of mechanical pump, and complicates
the angioplasty equipment and procedure.
With passive perfusion, the balloon catheter acts as a temporary stent.
Passive or auto perfusion catheters which have been proposed in the past
have used a design similar to "bail out" catheters: side holes in the
catheter through-lumen proximal and distal to the balloon. These
catheters, however, have several limitations. First, blood flow through
the balloon may be suboptimal for many clinical situations (such as distal
lesions and hypotension). Second, this configuration presents the problem
of catheter initiated thrombus, for which no satisfactory solution has
emerged. Third, the crossing profiles of these catheters typically require
two step angioplasty--the lesion is first opened with a low profile
catheter, which then must be exchanged with a perfusion balloon for a long
dilatation period.
SUMMARY OF THE INVENTION
The perfusion balloon catheter of the present invention includes a multiple
lumen shaft, a balloon head connected to a distal end of the shaft, a
balloon mounted around the balloon head, and a distal tip which extends
distally from the balloon. The balloon head defines a longitudinal blood
flow channel through which blood can flow when the balloon is inflated.
The blood flow channel is generally aligned with a guide wire lumen of the
shaft so that a guide wire can extend through the catheter and out the
distal tip, and can be partially withdrawn into the guide wire lumen to
permit perfusion through the blood flow channel while the balloon is
inflated.
BRIEF DESCRIPTION OF THE DRAWINGS
FIG. 1 shows a preferred embodiment of the passive perfusion balloon
catheter of the present invention.
FIG. 2 is a cross-sectional view of a distal portion of the passive
perfusion balloon catheter of FIG. 1.
FIG. 3 is a sectional view along Section 3--3 of FIG. 2.
FIG. 4 is a sectional view along Section 4--4 of FIG. 2.
FIGS. 5A, 5B, 5C are cross-sectional views of the passive perfusion balloon
catheter showing a guide wire extending through the balloon head and out
the distal tip, showing the guide wire in a retracted position to permit
perfusion through the balloon head, and showing the guide wire during
reinsertion through the balloon head, respectively.
FIG. 6 shows a sectional view of a distal portion of another embodiment of
the present invention.
FIG. 7 is a partial sectional view showing still another embodiment.
DETAILED DESCRIPTION OF THE PREFERRED EMBODIMENTS
Passive perfusion balloon catheter 10 shown in FIGS. 1 and 2 includes five
major elements: manifold 12, shaft 14, balloon head 16 (FIG. 2), balloon
18, and distal tip 20.
Manifold 12 includes inflation port 22, through port 23 and guide wire port
24. Hemostasis valve 26 is mounted on guide wire port 24. In FIG. 1, guide
wire 28 extends through guide wire port 24, shaft 14, balloon head 16, and
distal tip 20.
Shaft 14 is a multilumen tube. In preferred embodiments, shaft 14 is a dual
lumen polyethylene tube containing inflation lumen 30 and guide wire lumen
32. At its distal end, inflation lumen 30 is in communication with the
interior of balloon 18. Radiopaque fluid is supplied under pressure from
an inflation device (not shown) which is connected to inflation port 22 of
manifold 12.
Guide wire lumen 32 extends from guide wire port 24 to distal opening or
cut-away 34 which is positioned adjacent proximal opening 36 of balloon
head 16. Through port 23 communicates also with wire lumen 32.
Balloon head 16 forms a substrate or platform for balloon 18. Balloon head
16 defines a main blood flow channel which extends longitudinally through
balloon 18. Balloon head 16 is preferably formed by a polyolefin tube 40
which is reinforced by helically wound metal coil 42. As shown in FIG. 2,
coil 42 extends from the distal end of balloon head 16 to a position near
the proximal end of balloon head 16 to provide additional strength to
prevent collapse of balloon head tubing 40 due to the pressure of the
fluid supplied to the interior of balloon 18 during inflation. Coil 42
provides this additional collapse resistance without sacrificing
flexibility, and while decreasing kinkability. Coil 42 also acts as a
radiopaque marker. The use of coil 42 also permits a thin wall
construction to maximize the size of the flow passage without sacrificing
profile.
Proximal portion 44 of tubing 40 has a reduced profile and is shaped to
receive a distal portion of inflation lumen 30. As best shown in FIGS. 2
and 3, inflation lumen 30 extends beyond guide wire lumen 32 and has a
reduced profile in proximal bond area 45 where balloon head 16, inflation
lumen 30, and proximal portion 44 of tubing 40 are all bonded together.
This bond is formed by suitable adhesive, such as an epoxy.
Balloon 18, which is shown in its inflated state in FIGS. 1 and 2, is
bonded near its distal end to a distal end portion of balloon head 16. The
bond is formed, in preferred embodiments, by a suitable adhesive such as
epoxy.
Distal tip 20 is, in the embodiment shown in FIG. 2, an integral extension
of balloon 18. Tip 20 extends beyond the distal end of balloon head 16,
and is slightly conical with a distal opening 46 aligned with blood flow
channel 38. The purpose of distal tip 20 is to provide for atraumatic
tracking over guide wire 28.
In the preferred embodiment of the present invention, reinforcing coil 42
is formed of a helically wound flat ribbon of radiopaque metal such as a
stainless steel or a platinum alloy, and is bonded to the interior wall of
tubing 40. In order to provide a smooth interior surface for movement of
guide wire 28 and for blood flow, a coating of a thromboresistant material
is preferably formed on the interior of tubing 40 and coil 42.
FIGS. 5A-5C show guide wire 28 in three different positions, representing
three different stages in the use of passive perfusion balloon catheter
10. In FIG. 5A, guide wire 28 is shown extending through guide wire lumen
32, across the gap between distal opening 34 of shaft 14 and proximal
opening 36 of balloon head 16, through blood flow channel 38, and out
through distal opening 46 of distal tip 20.
In FIG. 5B, guide wire 28 has been withdrawn partially, so that the distal
end of guide wire 28 is within guide wire lumen 32. This opens up blood
flow passage 38 to allow perfusion of blood in a longitudinal direction
through balloon head 16 in a proximal to distal direction. The blood flow
channel 38 is sufficiently large so that balloon 18 can remain inflated
for prolonged periods of time without causing significant ischemia
distally from balloon 18.
FIG. 5C illustrates one of the important advantages of passive perfusion
balloon catheter 10 of the present invention. In FIG. 5C, guide wire 28 is
in the process of being reinserted through balloon head 16 and ultimately
out through distal opening 26 in distal tip 20. This allows an exchange
guide wire to be left in position across the stenosis after balloon 18 is
collapsed and catheter 10 is withdrawn.
The close spacing and alignment of guide wire lumen 32 and blood flow
passage 38 allows this reinsertion. Guide wire lumen 32 holds guide wire
28 in alignment, so that it can be reinserted if necessary through blood
flow passage 38.
In a preferred embodiment of the present invention, distal opening 34 at
the distal end of guide wire lumen 32 has a bevel angle of about 60
degrees, while proximal opening 36 at the proximal end of balloon head 16
forms a bevel angle of about 30 degrees. As illustrated in FIG. 2, the
proximal tip of balloon head 16 extends slightly into guide wire lumen 32.
The purpose of the bevelled openings 34 and 36 is to provide as
unobstructed a flow path as possible for blood entering blood flow channel
38. In addition, openings 34 and 36 provide a relatively smooth profile to
facilitate movement of catheter 10 during insertion and removal.
FIG. 6 shows a sectional view of the distal portion of a modified
embodiment of perfusion balloon catheter 10. In FIG. 6, the same reference
characters used in previous Figures are used to designate similar
elements. The catheter of FIG. 6 includes side port 100 in distal tip 20,
which facilitates blood flow even if distal opening 46 is blocked (for
example, because the artery has a sharp bend just beyond the stenosis and
opening 46 is positioned against an arterial wall.
FIG. 7 shows still another embodiment in which resilient flap 120 is
positioned adjacent opening 34 to urge guide wire 28 generally in a
direction toward inflation lumen 30. The force applied by flap 120 further
ensures that the distal tip of guide wire 28 will move out of opening 34
and through opening 36 into flow channel 38 during insertion (or
reinsertion) of guide wire 28.
Passive perfusion balloon catheter 10 of the present invention has several
advantages over other approaches to passive or auto perfusion. First, the
large diameter of blood flow channel 38 permits relatively high blood flow
rates through catheter 10 while balloon 18 is inflated. Second, flow
channel 38 is relatively short compared to the lumens of other devices,
which reduces the possibility of clotting. Third, because flow channel 38
is aligned with the primary flow axis of the artery, there is less trauma
to the blood, and less pressure head required for blood flow. Fourth, the
ability to remove guide wire 28 in order to permit perfusion, while
keeping it in a position to be reinserted and left across the stenosis
when catheter 10 is withdrawn offers an important option to the physician.
Although the present invention has been described with reference to
preferred embodiments, workers skilled in the art will recognize that
changes may be made in form and detail without departing from the spirit
and scope of the invention.
* * * * *
|
|
|
|
|
Description  |
|