|
Description  |
|
|
BACKGROUND OF THE INVENTION
The present invention relates to blood component testing generally, and
more particularly to blood glucose level testing devices.
Heretofore, various methods have been developed to carry out a program of
home glucose monitoring for diabetics. These methods typically require a
piercing device for piercing the skin of the patient to obtain a capillary
blood sample and a separate device or mechanism for testing the
concentration of glucose in the sample. Typically, automatic lancets,
which include a generally complex arrangement of elements that cooperate
to automatically trigger and retract a needle (lance), are used to pierce
the skin. During use, the tip of the automatic lancet is placed lightly
against the chosen site on a finger of a patient. A trigger is actuated
which causes the lance to penetrate the skin. Some of these lancets are
constructed such that the lance immediately retracts inside the lancet
after making the skin puncture. Then the automatic lancet is lifted away
from the finger and put aside, while a large drop of blood is allowed to
form at the puncture site. One way in which the blood sample is tested is
with what is known as a test strip. These test strips generally include
reagents which when placed in contact with the blood sample undergo a
change in color that indicates the glucose level of the sample. One such
test strip is made by Miles, Inc., Diagnostic Div., Elkhart, Ind. and
marketed under the name Glucofilm.TM.. These strips are typically stored
in a bottle to avoid contamination.
When removing a test strip from its bottle, the user must be careful not to
use the hand which has been pierced with the lancet. Otherwise, the blood
sample could be lost from the pierced finger. The strip is placed on a
flat dry surface with the target area of the strip (i.e., the reagent
area) facing upward. The punctured finger is then positioned such that the
blood forms a hanging drop which will naturally spread over the target
area on contact. A proper size drop forms a small dome that completely
covers the target. It is also important that the user not smear the blood
sample onto the test strip or cover the area of the test strip that
surrounds the reagent with the blood sample. Otherwise, inaccurate blood
glucose level readings can result.
One of the drawbacks of the procedure described above is that it is
awkward. Once the lance has been actuated to puncture a chosen site on the
finger, only one hand is available to put aside the automatic lancet,
obtain a test strip and position that strip such that a blood sample can
be properly deposited thereon.
In addition, the number of component elements used to carry out the method
creates inventory problems. For example, the user must maintain a supply
of lances for the automatic lancet and a supply of test strips. If the
lancet or one of these supply items is unavailable, the patient's glucose
level cannot be monitored. This can happen by inadvertently forgetting to
pack the automatic lancet or an appropriate supply of lances or test
strips when preparing for overnight or extended travel.
Therefore, there is a need to provide a system for testing blood glucose
levels that permits the procedure to be easily and reliably carried out.
The system also should simplify inventory requirements.
SUMMARY OF THE INVENTION
The present invention is directed to a device for testing the concentration
of a blood component in a blood sample that avoids the problems and
disadvantages of the prior art. The invention generally accomplishes this
goal by providing a modular construction that integrates a reagent
carrying member and a lance. The reagent carrying member is adapted for
application against the skin of a patient and contains a reagent that
indicates the concentration of a blood component in a blood sample placed
in contact therewith. The lance, which is used to pierce the skin of a
patient, is coupled to the reagent carrying member through a mechanism
that translates the lance between a first position where the distal end of
the lance is positioned on one side of the pad and a second position where
the distal end of the lance is positioned on the other side of the reagent
carrying member for puncturing the patient's skin.
The integrated construction eliminates the need for complicated and
relatively expensive automatic lancets, loading and unloading lancets with
replacement lance darts, and maintaining a separate supply of test strips
for testing the blood sample. In addition, with the integrated lance and
reagent carrying member construction, blood flows directly from the finger
to the reagent carrier, thereby eliminating complications that can arise
with conventional methods of placing a drop of the blood sample on a test
strip.
Another feature of the invention is that a plurality of the modular
sampler-testers can be interconnected in series to form a roll or strip of
testers. The number of modules in a given strip can be selected to provide
the user with a supply of testers sufficient for two weeks, a month, etc.
A modular tester is simply torn from the roll when the blood is to be
tested.
The above is a brief description of some deficiencies in the prior art and
advantages of the present invention. Other features, advantages and
embodiments of the invention will be apparent to those skilled in the art
from the following description, accompanying drawings and appended claims.
BRIEF DESCRIPTION OF THE DRAWINGS
FIG. 1 is a top plan view of the blood sampler and component tester in
accordance with the principles of the present invention;
FIG. 2 is a side elevational view of the invention illustrated in FIG. 1;
FIG. 3 is a longitudinal cross section of the invention illustrated in FIG.
1 showing the lance in the retracted state;
FIG. 4 is a longitudinal cross section of the invention illustrated in FIG.
1 showing the lance in the extended state;
FIG. 5 is a bottom plan view of the invention illustrated in FIG. 1, while
FIGS. 5a and 5b are views taken along line 5--5 showing a first and second
embodiment of the invention;
FIG. 6 is a bottom plan view of another embodiment of the invention
illustrated in FIG. 1; and
FIG. 7 is a bottom plan view of a further embodiment of the invention
illustrated in FIG. 1.
DESCRIPTION OF THE PREFERRED EMBODIMENTS
Referring to the drawings in detail, wherein like numerals indicate like
elements, the blood sampler and component tester is illustrated in
accordance with the principles of the present invention. Although the
sampler and tester will be described with respect to blood glucose
testing, the invention can be constructed to test other blood components
as will be apparent from the following description, drawings and appended
claims.
FIGS. 1 and 2 illustrate the modular construction of each integrated blood
sampler and tester. Referring to FIG. 3, each module 2 includes a
dome-shaped member 4 having a piercing element or lance 6 which is
arranged to pass through substrate 8 and reagent strip 10 to puncture the
skin of a patient's finger 12 such that a sampling of blood can be made
(FIG. 4). Dome-shaped member 4 includes a cup-shaped or convex portion 14
and a generally flat or planar portion 16. Piercing element or lance 6
extends from the central region of cup-shaped portion 14 and is preferably
integrally formed therewith, for example, by molding.
Dome-shaped member 4 is attached to generally planar substrate 8 by
securing generally planar portion 16 of the dome-shaped member to one side
of the substrate. This can be done in a way conventional to those skilled
in the art. For example, generally flat portion 16 of the dome-shaped
member can be adhesively secured, ultrasonically welded or fused with heat
to substrate 8. During manufacture, a plurality of dome-shaped members 4
can be molded to form a plurality of serially connected domes 4 which can
then be attached in mass to an appropriate length of substrate.
Referring to FIGS. 3 and 4, substrate 8 has a central opening 18 for
permitting lance 6 to readily pass through the substrate, which preferably
is plastic. Substrate 8 further includes a guide member 20 for guiding
lance 6 through opening 18 in substrate 8. Guide member 20 is illustrated
in the form of a hub member that is integrally formed with the substrate
such that central aperture 22 in hub member 20 is aligned with central
opening 18 in the substrate. Reagent strip 10, preferably comprising
paper, also can be provided with a hole to enhance the smooth passage of
lance 6 therethrough.
Hub member 20 also functions as a stop mechanism for limiting rectilinear
translation of lance 6. It is important that the lance does not extend
beyond reagent strip 10 and, thus, into the patient's finger more than
about 0.06 inches. One way in which the extension of lance 6 is limited is
illustrated in FIG. 4. In that figure, the lateral dimension of the
central aperture 22 of hub member 20 is less than the lateral dimension of
lance 6 in the region adjacent to the interior surface of cup-shaped
member 14. It should be understood that other ways to limit the
translation and, thus, the penetration of lance 6 can be used without
departing from the scope of the invention.
Referring to FIGS. 4 and 5, reagent strip 10 includes generally planar
carrier 10a and reagent pad 10b. One surface 30 of carrier 10a is
adhesively or mechanically attached to the side of substrate 8 opposite
dome-shaped member 4. The other surface, i.e., the exposed surface of
carrier 10a supports reagent containing pad 10b which is attached (e.g.,
adhesively or mechanically) to the exposed surface of carrier 10a (FIG.
5a). Reagent pad 10b contains a reagent that indicates the concentration
of glucose in a blood sample placed in contact therewith. When the reagent
pad is placed in contact with a blood sample, the reagent pad changes
color or hue according to glucose concentration in the blood sample.
Accordingly, the pad and strip composition and construction can be similar
to commercially available test strips for blood glucose testing such as
Glucofilm.TM. strips discussed above. Such test strips use standard
enzymes such as peroxidase (horseradish) and glucose oxidase, and standard
color enzyme substrates such as tetramethylbenzidine. Generally, the
glucose oxidase catalyzes the oxidation of glucose in the blood sample,
producing gluconic acid and hydrogen peroxide. In the presence of
peroxidase, the hydrogen peroxide oxidizes the tetramethylbenzidine
turning the test pad blue with an intensity proportional to the glucose
concentration. Other reagents can be used to produce other color changes
as is conventional to those of ordinary skill in the art.
Alternatively, the reagent pad can be embedded or encapsulated in reagent
strip 10 (FIG. 5b). As evident from FIG. 5b, reagent pad 10b extends the
entire thickness of carrier 10a. This construction is advantageous for
reasons discussed below. It also should be noted that carrier 10a can be
impregnated with reagent. In this way, the reagent also can be applied
such that it is present throughout the entire thickness of the carrier.
A color code is provided to interpret the meaning of the reagent color
change. Thus, the color code can include regions of different hue for a
particular color. Referring to FIG. 1, segments 24a, 24b and 24c can, for
example, represent light, medium and dark shades of a particular color,
wherein these shades indicate low, normal and high glucose levels. With
this arrangement, when the reagent matches light segment 24a in the color
code, the glucose level is low and insulin is needed. But when the reagent
matches the relatively dark segment 24c of the color code, the glucose
level is high which indicates that the diet of the patient should be more
carefully monitored.
The color code is shown on the top marginal surface of generally flat
portion 16. This position is especially advantageous when dome-shaped
member 4 and substrate 8 comprise transparent material such as clear
plastic. When this construction is used in conjunction with a reagent pad
that extends the entire thickness of the reagent strip (e.g., a pad
embedded in the strip as illustrated in FIG. 5), the user can
simultaneously view and compare the reagent color and color code.
Alternatively, the color code can be provided on the substrate beneath
transparent convex portion 14 of dome-shaped member 4. However, when
dome-shaped member 4 and substrate 8 are not transparent, the color
preferably is attached to the exposed undersurface of reagent strip 10.
FIG. 2 illustrates a strip 25 of serially connected blood sampler and
testing modules 2. The strip contains a supply of modules sufficient for
monitoring blood glucose level over the selected period of time. A
mechanically weakened portion, tear line 26, extends in the transverse
direction of the strip between adjacent modules so that the modules may be
separated for use. In this way, the user can simply tear off the terminal
or desired number of modules from the strip along the tear line. Although
the tear line is shown in FIG. 2 in the form of a groove that extends
through dome-shaped member 4, substrate 8 and reagent strip 10, other
conventional techniques can be used to provide a mechanically weakened
portion that facilitates separation of the modules. For example, a
perforated region that extends transversely between adjacent modules can
be used in lieu of tear line 26.
The method for using the disposable blood sampler and component tester will
be described with reference to FIGS. 3 and 4. First, the user simply tears
off a blood sampler and tester module 2 along tear line 26 to separate a
module from the supply strip. Then sterile strip 28, which seals and
protects reagent strip 10 from contamination, is peeled off. Sterile strip
28 is releasably secured to reagent strip 10 by adhesive and can be in the
form of conventionally known protective strips such as the protective
strips used on Band-Aids.RTM. which are bandages manufactured by Johnson &
Johnson. Once the protective strip 28 has been removed, module 2 is placed
on the patient's finger such that the reagent strip 10 is in contact with
the patient's skin. The convex portion of dome-shaped member 4 is
depressed, thereby causing lance 6 to slide through guide 20 and central
opening 18 in substrate 8, and puncture the skin of the patient (FIG. 3).
As soon as the puncture is made, the patient immediately lifts the finger
that was used to actuate the device (FIG. 4). The dome of dome-shaped
member 4 immediately pops back to its original position illustrated in
FIG. 3, thereby rapidly retracting lance 3 from the patient's finger. The
configuration of dome-shaped member 4, together with its thickness and
composition ensures that the dome portion 14 will rapidly spring back to
its original position. In other words, the construction of dome-shaped
member 4 and its connection with substrate 8 provides the dome-shaped
member with a natural bias that urges lance 6 toward the retracted state.
It has been found that a dome comprising polypropylene and having a
diameter of about 0.60-1.00 inches, a wall thickness of about 0.060-0.080
inches and a radius of curvature of about 1.20 inches provides the desired
results. After the blood has sufficiently contacted the reagent pad
through capillary flow, the user can then compare the color or hue of the
reagent pad to the color code to the glucose level of the blood sample.
The modular sampler and tester can be discarded after use.
Electronic sensors can be used to provide a more accurate reading of the
glucose concentration in the blood sample. Such sensors, which
electronically measure the glucose level, are commercially available such
as the ExacTech.TM. blood glucose sensor distributed by Baxter Travenol
Laboratories, Inc. These devices enable those who are visually impaired,
such as the color blind, to carry out glucose monitoring at home. However,
these sensors are relatively expensive and can be misplaced or lost.
Further embodiments of the present invention which are illustrated in
FIGS. 6 and 7, provide a simple and relatively inexpensive mechanism
integrated with the module to facilitate blood glucose monitoring by those
with certain visual impairments such as color blindness.
Referring to FIG. 6, a plurality of reagent regions or pads are arranged
linearly. Each reagent pad 30, 32, 34 and 36 is doped with different
reagent concentrations. By varying the concentration of the enzymes, or
more preferably the color enzyme substrate, one can alter the color
density for a given reaction. For example, pad 30 can contain a reagent
concentration that undergoes a sharp change in hue when exposed to a low
concentration of glucose in a blood sample. Pads 32, 34 and 36 can be
doped with reagent concentrations such that these pads sequentially
undergo a sharp change in hue as the concentration of glucose in the blood
sample increases. Since each sharp change in hue effectively corresponds
to an on-off signal, pads 30-36 can be doped to indicate to an individual
who is color blind whether his/her glucose level is in a range below
normal, normal or above normal. Furthermore, by increasing the number of
pads to a number greater than three (the number of ranges), greater
precision in measuring glucose concentration can be achieved. For example,
when pad 30 undergoes a sharp change in hue (on-position) and the
remaining pads do not undergo any change in hue (off-position), an
extremely low glucose level could be indicated. On the other hand, when
both pads 30 and 32 undergo a sharp change in hue and pads 34 and 36
remain unchanged (the off-position), the code could represent a sensed
glucose level that is very close to normal.
Although one arrangement is described above with reference to FIG. 6, it
should be understood that any other arrangement can be used which
incorporates the above-discussed principles. By way of example, FIG. 7
illustrates a further embodiment of the invention showing another
arrangement of reagent pads 38, 40, 42, 44, 46 and 48. Again, these pads
are doped with different reagent concentrations such that different on-off
combinations of the pads indicate different concentrations of glucose in
the blood sample.
The above is a detailed description of a particular embodiment of the
invention. It is recognized that departures from the disclosed embodiment
may be within the scope of the invention and that obvious modifications
will occur to a person skilled in the art. The full scope of the invention
is set out in the claims that follow and in their equivalents.
Accordingly, the claims and specifications should not be construed to
unduly narrow the full scope of protection to which the invention is
entitled. For example, the term "reagent strip" is chosen to aid
disclosure rather than limit the invention, and use of this term is not
intended to limit the reagent carrier to a long or narrow piece. This term
is used herein to generically describe a reagent carrier of any
configuration.
* * * * *
|
|
|
|
|
Description  |
|